X-ray sources · ZR / ZRB Series
DTXR-8ZR
An 8 kW fixed-anode X-ray source with an RS-232 interface.
- Output power
- 8 kW
- Output voltage
- 40–125 kV
- Power input
- 220 V AC ±10%, 50 Hz
Specifications
Power & general parameters
| Output power | 8 kW |
|---|---|
| Power supply | 220 V AC ±10%, 50 Hz |
| Inverter frequency | 220 kHz ±5% |
| kV accuracy | ±4% |
| mA accuracy | ≤ ±5% (mA >25 mA, ms >5)≤ ±(10% +1 mA) (mA ≤25 mA) |
| Anode | Fixed anode |
Exposure parameters
| Exposure voltage | 40–125 kV, 1 kV step |
|---|---|
| Exposure current | 10–160 mA |
| Exposure time | 2–2,000 ms |
| mAs range | 0.2–200 mAs |
| Exposure time accuracy | ≤ ±(5% +0.2 ms) |
| mAs accuracy | ≤ ±(5% +0.2 mAs) |
X-ray tube parameters
| Supported tube power | 8 kW |
|---|---|
| Focal spot | 2.0 |
| Exposure duty cycle | 1:60 |
| Leakage radiation loading factors | 125 kV / 0.24 mA |
| Target angle | 15° |
| Tube inherent filtration | 0.65 mm Al / 75 kV |
| Monoblock filtration | 1.5 mm Al / 75 kV |
| Supported filament current | 160 mA |
Interfaces, modes & mechanics
| Communication interface | RS-232 |
|---|---|
| Dimensions | 526 × 240 × 210 mm |
| Weight | 17 kg |
| Exposure modes | kV–mAs / kV–mA–ms |