X-ray sources · ZR / ZRB Series
DTXR-5ZR
A fixed-anode X-ray source available in 5, 5.3 and 5.6 kW configurations.
- Output power
- 5 / 5.3 / 5.6 kW
- Output voltage
- 40–125 kV
- Power input
- 220–240 V AC
Specifications
Power & general parameters
| Output power | 5 / 5.3 / 5.6 kW |
|---|---|
| Power supply | 220–240 V AC |
| Inverter frequency | 100 kHz ±10% |
| kV accuracy | ±10% |
| Anode | Stationary anode |
| mA accuracy | ±10% |
Exposure parameters
| Exposure voltage | 40–125 kV, 1 kV step |
|---|---|
| Exposure current | 10–100 mA |
| Exposure time | 1–4,000 ms |
| mAs range | 0.1–320 mAs |
| Exposure time accuracy | ≤ ±(5% +0.2 ms) |
| mAs accuracy | ≤ ±(5% +0.2 mAs) |
X-ray tube parameters
| Supported tube power | 5 / 5.3 kW configurations: 5.3 kW5.6 kW configuration: 5.6 kW |
|---|---|
| Focal spot | Small 0.6large 1.8 |
| Exposure duty cycle | 1:60 |
| Supported small-focus current | 20 mA |
| Supported large-focus current | 100 mA |
| Leakage radiation loading factors | 125 kV / 0.24 mA |
| Target angle | 15° |
| Tube inherent filtration | 0.6 mm Al / 75 kV |
| Monoblock filtration | 1.5 mm Al / 75 kV |
Interfaces, modes & mechanics
| Communication interface | RS-232 |
|---|---|
| Exposure modes | kV–mAs / kV–mA–ms |