X-ray sources · ZR / ZRB Series

DTXR-5ZR

A fixed-anode X-ray source available in 5, 5.3 and 5.6 kW configurations.

Output power
5 / 5.3 / 5.6 kW
Output voltage
40–125 kV
Power input
220–240 V AC

Specifications

Power & general parameters
DTXR-5ZR · Power & general parameters
Output power5 / 5.3 / 5.6 kW
Power supply220–240 V AC
Inverter frequency100 kHz ±10%
kV accuracy±10%
AnodeStationary anode
mA accuracy±10%
Exposure parameters
DTXR-5ZR · Exposure parameters
Exposure voltage40–125 kV, 1 kV step
Exposure current10–100 mA
Exposure time1–4,000 ms
mAs range0.1–320 mAs
Exposure time accuracy≤ ±(5% +0.2 ms)
mAs accuracy≤ ±(5% +0.2 mAs)
X-ray tube parameters
DTXR-5ZR · X-ray tube parameters
Supported tube power5 / 5.3 kW configurations: 5.3 kW5.6 kW configuration: 5.6 kW
Focal spotSmall 0.6large 1.8
Exposure duty cycle1:60
Supported small-focus current20 mA
Supported large-focus current100 mA
Leakage radiation loading factors125 kV / 0.24 mA
Target angle15°
Tube inherent filtration0.6 mm Al / 75 kV
Monoblock filtration1.5 mm Al / 75 kV
Interfaces, modes & mechanics
DTXR-5ZR · Interfaces, modes & mechanics
Communication interfaceRS-232
Exposure modeskV–mAs / kV–mA–ms

DTXR-5ZR

DTXR-5ZR