X-ray sources · ZRFM / ZRF Series

DTXR-517ZRF

A rotating-anode source with 5 kW continuous fluoroscopy and 17 kW exposure and pulsed operation.

Output power
17 kW
Output voltage
40–130 kV
Power input
220 V AC ±10%, 50–60 Hz

Specifications

Power & general parameters
DTXR-517ZRF · Power & general parameters
Output powerContinuous fluoroscopy 5 kWexposure / pulsed fluoroscopy 17 kW
Power supply220 V AC ±10%, 50–60 Hz
Inverter frequency500 kHz ±5%
kV accuracy±4%
mA accuracy≤ ±5% (mA >25 mA, ms >5)≤ ±(10% +1 mA) (mA ≤25 mA)
AnodeRotating anode
Exposure parameters
DTXR-517ZRF · Exposure parameters
Exposure voltage40–130 kV, 1 kV step
Exposure current10–250 mA
Exposure time1–6,300 ms
mAs range0.1–320 mAs
Exposure time accuracy≤ ±(5% +0.2 ms)
mAs accuracy≤ ±(5% +0.2 mAs)
Fluoroscopy parameters
DTXR-517ZRF · Fluoroscopy parameters
Fluoroscopy voltage40–125 kV
Continuous fluoroscopy current0.5–6 mA
Pulsed fluoroscopy current10–40 mA, accuracy 10%
Fluoroscopy frame rate1–30 F/s
Pulse widthSupports 0.1 s, depending on frame rate
X-ray tube parameters
DTXR-517ZRF · X-ray tube parameters
Supported tube power17 kW
Focal spotSmall 0.3large 0.6
Exposure duty cycle1:60
Supported small-focus current100 mA
Supported large-focus current350 mA
Anode heat capacity150 kJ (200 kHU)
Interfaces, modes & mechanics
DTXR-517ZRF · Interfaces, modes & mechanics
Communication interfaceRS-232
DimensionsControl box 410 × 310 × 410 mmmonoblock 342 × 180 × 210 mm
WeightControl box 26 kgmonoblock 20 kg
Exposure modeskV–mAs / kV–mA–ms

DTXR-517ZRF

DTXR-517ZRF