DR · HFD Series · Dual tube

DTXR-50HFD

50 kW DR high-voltage generator.

Output power
50 kW
Output voltage
40–150 kV
Power input
380 V AC ±10%, 50 Hz, 3 phase

Specifications

Power & general parameters
DTXR-50HFD · Power & general parameters
Output power50 kW
Power supply380 V AC ±10%, 50 Hz, 3 phase
Inverter frequency160–500 kHz ±5%
kV accuracy±4%
mA accuracy≤ ±5% (mA >25 mA, ms >5)≤ ±(10% +1 mA) (mA ≤25 mA)≤ ±20% (ms ≤5)
Rotor speedLow speed
Exposure parameters
DTXR-50HFD · Exposure parameters
Exposure voltage40–150 kV, 1 kV step
Exposure current10–630 mA
Exposure time1–6,300 ms
Exposure time accuracy≤ ±(5% +0.2 ms)
mAs range0.1–630 mAs
mAs accuracy≤ ±(5% +0.2 mAs)
Fluoroscopy parameters
DTXR-50HFD · Fluoroscopy parameters
Fluoroscopy voltage40–125 kV, 1 kV step
Continuous fluoroscopy current0.5–6 mA
Pulsed fluoroscopy current10–40 mA
Pulsed fluoroscopy frame rate1–30
AEC & auxiliary power
DTXR-50HFD · AEC & auxiliary power
AECSupported
Auxiliary power24 V AC / 8 A
Interfaces, modes & mechanics
DTXR-50HFD · Interfaces, modes & mechanics
Exposure modeskV–mA (AEC) / kV–mAs / kV–mA–ms
APR entriesSupports 2,000 entries
Dual tubeSupported
Communication interfaceRS-232
Dimensions741 × 407 × 653

DTXR-50HFD

DTXR-50HFD