CT · DTCT-L Series
DTCT-50L
50 kW CT high-voltage generator.
- Output power
- 50 kW
- Output voltage
- 60–140 kV
- Power input
- 400 V AC +15% / −20%, 47–63 Hz, 3 phase
Specifications
Power & general parameters
| Output power | 50 kW |
|---|---|
| Power supply | 400 V AC +15% / −20%, 47–63 Hz, 3 phase |
| kV accuracy | ±0.5% of set value |
| mA accuracy | 0.5% |
| kV repeatability | ±0.1% between two exposures at the same voltage, 60–140 kV, output power >2 kW |
| kV ripple | >10 kHz: 2% of anode-to-cathode RMS voltage<10 kHz: 0.2% |
| Rotor speed | Low speed 50 / 60 Hzhigh speed 150 / 180 Hz, according to the matched tube |
Exposure parameters
| Exposure voltage | 60–140 kVbipolar mode 30–70 kV |
|---|---|
| Exposure current | 10–420 mA |
| Exposure time | >42 kW & ≤50 kW: >60 s>36 kW & ≤42 kW: >80 s≤36 kW: >102 s |