X-ray sources · DTCB-25 / DTCBCT-10
DTCBCT-10
A rotating-anode X-ray source with 10 kW exposure and 5 kW fluoroscopy power.
- Output power
- 10 kW
- Output voltage
- 40–130 kV
- Power input
- 220 V AC ±10%, 50 Hz
Specifications
Power & general parameters
| Output power | Exposure 10 kWfluoroscopy 5 kW |
|---|---|
| Power supply | 220 V AC ±10%, 50 Hz |
| Inverter frequency | 500 kHz ±5% |
| kV accuracy | ±4% |
| Anode | Rotating anode |
| mA accuracy | ≤ ±5% (mA >25 mA, ms >5)≤ ±(10% +1 mA) (mA ≤25 mA) |
Exposure parameters
| Exposure voltage | 40–130 kV, 1 kV step |
|---|---|
| Exposure current | 10–200 mA |
| Exposure time | 1–6,300 ms |
| mAs range | 0.1–200 mAs |
| Exposure time accuracy | ≤ ±(5% +0.2 ms) |
| mAs accuracy | ≤ ±(5% +0.2 mAs) |
Fluoroscopy parameters
| Fluoroscopy voltage | 40–125 kV |
|---|---|
| Continuous fluoroscopy current | 0.5–6 mA, 0.1 mA step |
| Pulsed fluoroscopy current | 10–100 mA, 0.5 mA step, accuracy 10% |
| Fluoroscopy frame rate | 1–30, step 1 |
X-ray tube parameters
| Supported tube power | 16 kW |
|---|---|
| Focal spot | Small 0.3large 0.6 |
| Exposure duty cycle | 1:60 |
| Supported small-focus current | 104 mA |
| Supported large-focus current | 380 mA |
| Leakage radiation loading factors | 130 kV / 0.22 mA |
| Target angle | 16° |
| Anode heat capacity | 80 kJ (107 kHU) |
Interfaces, modes & mechanics
| Communication interface | RS-232 |
|---|---|
| Dimensions | Control box 437 × 335 × 239 mminverter 379 × 237 × 239 mmmonoblock 342 × 180 × 206 mm |