X-ray sources · DTCB-05 Series

DTCB-05R

A rotating-anode X-ray source with a monoblock and inverter for exposure and fluoroscopy.

Output power
5 kW
Output voltage
40–125 kV
Power input
220 V AC ±10%, 50 Hz

Specifications

Power & general parameters
DTCB-05R · Power & general parameters
Output powerExposure 5 kWpulsed fluoroscopy 5 kWcontinuous fluoroscopy 500 W
Power supply220 V AC ±10%, 50 Hz
Inverter frequency300 kHz ±5%
kV accuracy±4%
AnodeRotating anode
mA accuracy≤ ±5% (mA >25 mA, ms >5)≤ ±(10% +1 mA) (mA ≤25 mA)
Inherent filtration2.1 mm Al (tube 0.6 mm Almonoblock 1.5 mm Al)
Exposure parameters
DTCB-05R · Exposure parameters
Exposure voltage40–125 kV, 1 kV step
Exposure current10–100 mA
Exposure time2–4,000 ms
mAs range0.1–200 mAs
Exposure time accuracy≤ ±(5% +0.2 ms)
mAs accuracy≤ ±(5% +0.2 mAs)
Fluoroscopy parameters
DTCB-05R · Fluoroscopy parameters
Fluoroscopy voltage40–125 kV, 1 kV step
Continuous fluoroscopy current0.5–6 mA, 0.1 mA step
Pulsed fluoroscopy current10–50 mA, 0.5 mA step
Fluoroscopy frame rate1–30, step 1
X-ray tube parameters
DTCB-05R · X-ray tube parameters
Supported tube power17 kW
Focal spotSmall 0.3large 0.6
Exposure duty cycle1:60
Supported small-focus current100 mA
Supported large-focus current350 mA
Leakage radiation loading factors125 kV / 0.24 mA
Target materialTungsten
Anode heat capacity150 kJ (200 kHU)
Interfaces, modes & mechanics
DTCB-05R · Interfaces, modes & mechanics
Communication interfaceRS-232
DimensionsInverter 430 × 305 × 100 mmmonoblock 305 × 138 × 243 mm
WeightInverter 6 kgmonoblock 21 kg
Exposure modeskV–mAs / kV–mA–ms
APR entriesSupports 2,000 entries

DTCB-05R