X-ray sources · DTCB-05 Series

DTCB-05

A fixed-anode monoblock and inverter supporting exposure, pulsed fluoroscopy and continuous fluoroscopy.

Output power
5 kW
Output voltage
40–125 kV
Power input
220 V AC ±10%, 50 Hz

Specifications

Power & general parameters
DTCB-05 · Power & general parameters
Output powerExposure 5 kWpulsed fluoroscopy 5 kWcontinuous fluoroscopy 500 W
Rated supply220 V AC ±10%, 50 Hz
Inverter frequency220 kHz ±5%
kV accuracy±4%
mA accuracy≤ ±5% (mA >25 mA, ms >5)≤ ±(10% +1 mA) (mA ≤25 mA)
AnodeFixed anode
Inherent filtration2.1 mm Al (tube 0.6 mm Almonoblock 1.5 mm Al)
Exposure parameters
DTCB-05 · Exposure parameters
Exposure voltage40–125 kV, 1 kV step
Exposure current10–100 mA
Exposure time2–4,000 ms
Exposure time accuracy≤ ±(5% +0.2 ms)
mAs range0.1–200 mAs
mAs accuracy≤ ±(5% +0.2 mAs)
Fluoroscopy parameters
DTCB-05 · Fluoroscopy parameters
Fluoroscopy voltage40–125 kV, 1 kV step
Continuous fluoroscopy current0.5–6 mA, 0.1 mA step
Pulsed fluoroscopy current10–50 mA, 0.5 mA step
Frame rate1–30, step 1
X-ray tube parameters
DTCB-05 · X-ray tube parameters
Tube anode heat capacity53 kJ (74 kHU)
Interfaces, modes & mechanics
DTCB-05 · Interfaces, modes & mechanics
Exposure modeskV–mAs / kV–mA–ms
APRSupports 2,000 APR entries
Communication interfaceRS-232
Inverter dimensions / weight340 × 240 × 80 mm / 5.5 kg
Monoblock dimensions / weight310 × 130 × 210 mm / 11.5 kg

DTCB-05

DTCB-05